Yes, this is a new CAP requirement (as of April 2014) and is a phase 2 deficiency if it is not followed:
“The performance of new lots/shipments of antibody and detection system reagents is compared with old lots/shipments before or concurrently with being placed into service.
NOTE: Parallel staining is required to control for variables such as disparity in the lots of detection reagents or instrument function. Comparing the results of the new reagent run in parallel on the same fresh control (patient or normal) is recommended to ensure that the new reagent lot provides a clinically comparable result to the old reagent lot. Testing the new reagent on a standardized control with defined acceptance criteria is also acceptable. Before using a new lot of antibody cocktail, it is recommended to compare each individual antibody by using a side scatter vs. fluorescence plot and having defined acceptance criteria from old lot to new lot. Only when there is agreement between the individual old and new lots should the cocktail be made and then tested against the old cocktail.” (CAP Flow Cytometry Checklist 04/21/2014)
In most instances, a normal peripheral blood can be used to fulfill this requirement. In our laboratory, we routinely follow this process:

Further Reading:
1. ICSH/ICCS Workgroup (2013) Validation of cell-based fluorescence assays: Practice guidelines from the International Council for Standardization of Haematology and International Clinical Cytometry Society. Cytometry B Clin Cytom. 84:281.